Requirements

● 220 volts ± %5, 50Hz 3.5 Kw Single phase main electric output.

● Minimum 3 x 1.5mm Power cable line and 16 Amp fuse installation.

● Grounding type Class 3 ( Class 3 Grounding requirements are max 100 Ω grounding resistance and 500 V megger measuring instrument).

● Minimum P=6 bar compressed air (ISO 9001, ISO 14001 & ISO 1217).

● Minimum 100 L/min flow rate. Recommended 175 L/min flow rate.

● The air line from compressor to Machine must be at least 3/4” size.

● Manual valve for on/off.

● Solid impurities – Class 3 – Filter grade at least 5μm for solids

● Water content – Class 4 – max. Pressure dew point +3C

● Total oil content – Class 2 – max. Oil content 0.1mg/m3.

● The machine generates vibration while working therefore a smooth concrete floor surface should be prepared. Only a 5mm level difference is allowed throughout the machine base. The machine's surrounding should be clear.

● Between the wall and the machine, there should be a minimum distance of 50 cm.

● If the humidity and temperature of the compressed air are variable, then the drying system and air coolers should be installed.

● Compressed air temperature should be the same as the room temperature.

● The ambient temperature for machining conditions should be 20-25 °C degrees.

Cutting fluids can be found in our store.

To protect the machine from electricity blackouts, minimum 3,5 kVA Online UPS should be installed.

Atlas Copco GX2 Model air compressor is recommended.

Package information

D43N + table + liquid tank

Description: Yenadent D43N CNC Machining Center

Width dimension: 82 cm

Depth dimension: 92 cm

Height dimension: 181 cm

Gross weight: 330 kg

Net weight: 288 kg

Chiller unit

Width dimension: 67 cm

Depth dimension: 37 cm

Height dimension: 68 cm

Gross weight: 33 kg

Net weight: 23 kg

Options

Vacuum unit

Width dimension: 37 cm

Depth dimension: 47 cm

Height dimension: 59 cm

Gross weight: 19 kg

Net weight: 14 kg

Unboxing

Assembly

Tablet PC

Air hose

Outer diameter: 12mm
Inner diameter: 8mm

Machine power cable

Chiller unit

Chiller connection cable

Cover removal

User manual

USER MANUAL PDF'İ

Product added to wishlist
Product added to compare.